Liquiline System CA80NO
Maintenance
Hauser
147
4. Once the action is completed, select
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Rinse analyzer
must be selected.
5. Immerse all the hoses (RK, S1, P) in a beaker containing approx. 200 ml (6.76 fl.oz)
of distilled water in order to rinse the system with water.
6. Select
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Sample collector
/
Empty sample collector
/
Start
must be selected.
The system is automatically rinsed with sample and emptied. This process takes
approx. 4 minutes.
7. Remove the hoses from the beaker and dry them with a clean paper towel.
8. Select
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Rinse analyzer
once
again to rinse the hoses with air.
9. The analyzer can now be disconnected from the mains supply.