Liquiline System CA80NO
Diagnostics and troubleshooting
Hauser
119
11.5 Adapting the diagnostic information
11.5.1 Classification of diagnostic messages
In the
DIAG
/
Diagnostics list
menu you can find more detailed information on the current
diagnostic messages displayed.
In accordance with Namur specification NE 107, the diagnostic messages are characterized
by:
• Message number
• Error category (letter in front of the message number)
•
F
= (Failure) a malfunction has been detected
The measured value of the affected channel is no longer reliable. The cause of the
malfunction is to be found in the measuring point. Any control system connected
should be set to manual mode.
•
C
= (Function check), (no error)
Maintenance work is being performed on the device. Wait until the work has been
completed.
•
S
= (Out of specification), the measuring point is being operated outside its
specification
Operation is still possible. However, you run the risk of increased wear, shorter
operating life or lower measurement accuracy. The cause of the problem is to be found
outside the measuring point.
•
M
= (Maintenance required), action should be taken as soon as possible
The device still measures correctly. Immediate measures are not necessary. However,
proper maintenance efforts would prevent a possible malfunction in the future.
• Message text
If you contact the Service Department, please cite the message number only. Since you
can individually change the assignment of an error to an error category, the Service
Department cannot use this information.
11.5.2 Adapting the diagnostic behavior
All the diagnostic messages are assigned to specific error categories at the factory. Since
other settings might be preferred depending on the application, error categories and the
effect errors have on the measuring point can be configured individually. Furthermore,
every diagnostic message can be disabled.
Example
Diagnostic message 531
Logbook full
appears on the display. You want to change this
message so that an error is not shown on the display for example.
1. For analyzer-specific diagnostic messages, select
Menu
/
Setup
/
Analyzer
/
Extended
setup
/
Diagnostics settings
/
Diag. behavior
and for sensor-related messages, select
Menu
/
Setup
/
Inputs
/<Sensor>/
Extended setup
/
Diagnostics settings
/
Diag.
behavior
.
2. Select the diagnostics message and press the navigator button.
3. Decide: (a) Should the message be deactivated? (
Diagnostics message
=
Off
)
(b) Do you want to change the error category?(
Status signal
)
(c) Should an error current be output? (
Failure current
=
On
)
(d) Do you want to trigger a cleaning program? (
Cleaning program
)
4. Example: You deactivate the message.
The message is no longer displayed. In the
DIAG
menu, the message appears as
Past message
.
Possible settings
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.