Operation
Liquiline System CA80NO
94
Hauser
In addition
you can define 8 digital device variables (DI):
‣
Define the data source.
You can choose from relays, binary inputs and limit switches.
For further information on "EtherNet/IP", see the Guideline for Communication via
EtherNet/IP, SD01293C
10.8 Additional functions
10.8.1 Limit contactors
There are different ways to configure a limit switch:
• Assigning a switch-on and switch-off point
• Assigning a switch-on and switch-off delay for a relay
• Setting an alarm threshold and also outputting an error message
• Starting a cleaning function (for sensors)
Menu/Setup/Additional functions/Limit switches/Limit switch 1 ... 8
Function
Options
Info
Source of data
Selection
• None
• Sensor inputs
• Binary inputs
• Controller
• Fieldbus signals
• Mathematical functions
• MRS set 1 ... 2
Factory setting
None
‣
Specify the input or output which is to be the
data source for the limit switch.
The sources of data on offer depend on your
device version.
You can choose from connected sensors, binary
inputs, fieldbus signals, mathematical functions,
controllers and sets for measuring range
switching.
Measuring value
Selection
Depends on:
Source of data
‣
Select the measured value, see the following
table.
Measured value
depending on the
Source of data
Source of data
Measured value
Nitrite
Selection
Main value
pH Glass
Selection
• Raw value mV
• pH
• Temperature
pH ISFET
ORP
Selection
• Temperature
• ORP mV
• ORP %
Oxygen (amp.)
Selection
• Temperature
• Partial pressure
• Concentration liquid
• Saturation
• Raw value nA
(only
Oxygen (amp.)
)
• Raw value μs
(only
Oxygen (opt.)
)
Oxygen (opt.)