Modbus RS485
Signal encoding
EIA/TIA-485
Data transmission rate
2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200
baud
Galvanic isolation
Yes
Bus termination
Internal slide switch with LED display
Web server and Modbus TCP
Signal encoding
IEEE 802.3 (Ethernet)
Data transmission rate
10 / 100 MBd
Galvanic isolation
Yes
Connection
RJ45, M12 optional
IP address
DHCP or configuration using menu
EtherNet/IP
Signal encoding
IEEE 802.3 (Ethernet)
Data transmission rate
10 / 100 MBd
Galvanic isolation
Yes
Connection
RJ45, M12 optional (D-encoded)
IP address
DHCP (default) or configuration via menu
Signal on alarm
Adjustable, as per NAMUR Recommendation NE 43
• In measuring range 0 to 20 mA:
Failure current from 0 to 23 mA
• In measuring range 4 to 20 mA:
Failure current from 2.4 to 23 mA
• Factory setting for failure current for both measuring ranges:
21.5 mA
Load
Max. 500 Ω
Transmission behavior
Linear
15.3 Current outputs, active
Span
0 to 23 mA
Signal characteristic
Linear
Electrical specification
Output voltage
Max. 24 V
Test voltage
500 V