Liquiline System CA80NO
Diagnostics and troubleshooting
Hauser
135
DIAG/System information
Function
Options
Info
Save system information
Save to SD card
File name assigned
automatically (includes a
time stamp)
The information is saved on the SD
card in a "sysinfo" subfolder.
The csv file can be read and edited
in MS Excel, for example. This file
can be used when servicing the
device.
1)
Provided you give the manufacturer all the information about changes to the hardware.
11.11 Sensor information
‣
Select the channel you want from the list of channels.
Information in the following categories is displayed:
•
Extreme values
Extreme conditions to which the sensor was previously exposed, e. g. min./max.
temperatures
•
Operating time
Operating time of the sensor under defined extreme conditions
•
Calibration information
Calibration data of the last calibration
•
Sensor specifications
Measuring range limits for main measured value and temperature
•
General information
Information on sensor identification
The specific data that are displayed depends on what sensor is connected.
11.12 Simulation
You can simulate values at inputs and outputs for testing purposes:
• Current values at current outputs
• Measured values at inputs
• Relay contact opening or closing
Only current values are simulated. Via the simulation function, it is not possible to
calculate the totalized value for the flow or rainfall.
‣
Before simulation: Enable the inputs and outputs in the Setup menu.
DIAG/Simulation
Function
Options
Info
Current output x:y
Simulation of an output current
This menu appears once for each current output.
Simulation
Selection
• Off
• On
Factory setting
Off
If you simulate the value at the current output,
this is indicated on the display by a simulation
icon in front of the current value.
Current
2.4 to 23.0 mA
Factory setting
4 mA
‣
Set the desired simulation value.
3)
Not available for all sensor types.