Liquiline System CA80NO
Electrical connection
Hauser
37
Example: Connecting the cleaning unit 71072583 for CAS40D
NOTICE
Power consumption too high for the Liquiline alarm relay!
Can cause irreparable damage to the base module
‣
Connect the cleaning unit only to terminals of an additional module (AOR, 2R or 4R),
not
to the alarm relay of the base module.
Example: Connecting the Chemoclean CYR10 injector cleaning unit
41
43
42
Relay 1
41
43
42
Relay 2
1
2
1
2
N
L
PE
CM44x/CM44xR/
CM44P
CYR10
1
2
3
4
5
A0028598
51
Connecting the CYR10 injector cleaning unit
1
External power supply
2
Cleaner to spray head
3
Container with cleaner
4
Motive water 2 to 12 bar (30 to 180 psi)
5
Backflow valve (to be provided by the customer)
6.4.4
Connecting digital communication
Module 485
52
Module
Ethernet
81
82
DIP switch
on
1
2
4
8
16
32
64
128/SW
Service
T
COM
SF
BF
PWR
DGND
VP
96'
95'
99'
Termi-
nation
DP/
RS485
96
95
99
53
Wiring
diagram