Electrical connection
Liquiline System CA80NO
30
Hauser
5. Comply with the instructions for protective grounding when installing the analyzer.
6. Remove the protective cover in the upper right-hand corner.
A0028925
7. Connect the cleaning valve to the following plug-in terminals:
SP1
N
L
PE
86
85
97
98
BK
BK
GN/YE
GY
PK
GN
YE
1
2
A0028926
29
Connection scheme for Liquiline System CAT810
1
Liquiline System CAT810, 100 to 120 V/200 to 240 V AC
2
Are not used
8. Secure the protective cover after connecting. Make sure that no cables or hoses are
jammed.
9. Use the 6 screws to secure the carrier board after connecting.
6.3.2
Connecting the optional hose heater and communication
between the CAT820/CAT860 and the analyzer
Only use terminated original cables where possible. The sensor cable, fieldbus cable
and Ethernet cable must be shielded cables.
Cable sample (does not necessarily correspond to the original cable supplied)
1
2
3
30
Terminated cable
1
Outer shield (exposed)
2
Cable cores with ferrules
3
Cable sheath (insulation)
SP1
N
L
PE
86
85
97
98
4
5
31
Inserting the cable
4
Shielding clamp for Memosens
supply and power supply
5
Cable clamp for hose heating
SP1
N
L
PE
86
85
97
98
32
Tighten the screw (2 Nm)
1. Disconnect the mains plug.