Operation
Liquiline System CA80NO
70
Hauser
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Selection
100 to 1000 ml
Factory setting
Reagent RK
Selection
100 to 1000 ml
Factory setting
1000 ml
Warning limits
Standard S1
Selection
1 to 20 %
Factory setting
2 %
Reagent RK
Selection
1 to 40 %
Factory setting
10 %
Diag. code 726
Read only
Alarm limits
Standard S1
Selection
1 to 20 %
Factory setting
2 %
Reagent RK
Selection
1 to 40 %
Factory setting
5 %
Absorption curve recording
In automatic mode
Selection
• Off
• On
Factory setting
Off
Off
: Recording is only possible in the manual
mode
On
: Recording also takes place in the automatic
mode
Curve
Selection
1 to 7
Factory setting
2
Select the absorption curve to be recorded.
Only one curve can be selected at any one time.
The data are saved in a logbook.
After power failure
Selection
• Last mode
• Manual mode
Factory setting
Last mode
Setting specifying how the analyzer should
behave after a power failure, or when the power
is back on.
Last mode
: The analyzer remains in the mode
that was last set. Example: The automatic mode
was set. The analyzer continues after
initialization and after any samples have been
discarded.
Manual mode
: The analyzer switches to the
manual mode and waits for user action.
1)
"%C" stands for context-dependent text that is automatically generated by the software. This then contains
the name of the sample preparation, e.g. "1" or "2".