Liquiline System CA80NO
Operation
Hauser
65
Menu/Setup/General settings/Extended setup/Ethernet
Function
Options
Info
Netmask
xxx.xxx.xxx.xxx
On the basis of the IP address of a device, the
netmask specifies which IP addresses this device
searches for in its own network and which
addresses it could access in other networks via a
router. It therefore divides the IP address into a
network part (network prefix) and a device part.
The network part must be identical for all
devices in the individual network, and the device
part must be different for every device within the
network.
Gateway
x.x.x.x
A gateway (protocol converter) enables
communication between networks that are
based on completely different protocols.
Service switch
Read only
MAC-Address
Read only
The MAC address (Media Access Control
address) is the hardware address of every
individual network adapter which is used to
uniquely identify the device in a computer
network.
EtherNetIP Port 44818
Read only
A port is a part of an address which assigns data
segments to a network protocol.
Accepting settings
Have you changed settings manually, such as the IP address?
‣
Before you leave the
Ethernet
menu:
SAVE
Select , to apply your settings.
In the
DIAG
/
System information
menu you can check whether the new settings
are used.
Data management
Firmware update
Please contact your local sales office for information on firmware updates available
for your controller and its compatibility with earlier versions.
Current firmware version
of the analyzer, actuator control module, photometer and
sample preparation 1 :
Menu
/
Diagnostics
/
System information
/.
‣
Back up your current setup and your logbooks to an SD card.
To install a firmware update, you must have the update available on an SD card.
1. Insert the SD card into the controller card reader.
2. Go to
Menu
/
Setup
/
General settings
/
Extended setup
/
Data management
/
Firmware update
.
The update files on the SD card are displayed.
3. Select the desired update and select yes when asked the following:
The current firmware will be overwritten.
After this the device will reboot.
Do you want to proceed?
The firmware is loaded and the device is then started with the new firmware.