Operation
Liquiline System CA80NO
54
Hauser
10
Operation
10.1 Display
10.1.1 Soft keys in the measurement mode
On the bottom row of the display you can find four soft keys in the measuring screens:
• With
MENU
and
DIAG
take you directly to the specific software menu.
• With
HOLD
you can activate an immediate, general hold for sensors. This also sets all
the linked outputs, controllers and cleaning cycles to HOLD. Any sensor cleaning
programs that are currently running are interrupted. However, you can also start manual
sensor cleaning when a hold is active.
• With
MODE
takes you to a picklist of frequently used software functions.
10.1.2 Measuring mode
There are different display modes: (press the navigator button to change the mode)
• (1) Overview of all the inputs and outputs
• (2) Analyzer measured value or main measured value of an input or output or the status
of a relay
• (3) Smaller visualization of the analyzer measured value or main and secondary
measured value in the case of a sensor input
• (4) Analyzer measured value and pending actions or all measured values in the case of a
sensor input
• (5)
Only for analyzer measured value:
Graphic display
There are also submenus:
(6) User-definable measuring menus (only available if already defined)
Choice of previously configured screens
Change of channel to modes (2) - (5)
‣
Turn the navigator.
The display changes from channel to channel.
10.1.3 Device status
Icons on the display alert you to special device states.
Icon
Location
Description
Header bar
Diagnostic message "Failure"
Header bar
Diagnostic message "Maintenance request"
Header bar
Diagnostic message "Check"
Header bar
Diagnostic message "Out of specification"
Header bar
Fieldbus or TCP/IP communication active
Header bar
Hold active (for sensors)
At measured value
Hold for the actuator (current output, limit switch etc.) is active
At measured value
An offset has been added to the measured value
At measured value
Measured value in "Bad" or "Alarm" state
At measured value
Automatic temperature compensation active (for sensors)
At measured value
Manual temperature compensation active (for sensors)
Header bar
Simulation mode active or Memocheck SIM connected