Operation
Liquiline System CA80NO
112
Hauser
Menu/Setup/Additional functions/Mathematical functions/MF1 to 6/Mode = Degassed conductivity
Function
Options
Info
Calculation
Selection
• Off
• On
Factory setting
Off
On/off switch for the function
Cation conductivity
Connected conductivity
sensor
Cation conductivity
is the sensor downstream
from the cation exchanger and upstream from
the "degassing module",
Degassed conductivity
is the sensor at the outlet of the degassing
module.
Measured value interrogation is obsolete as you
can only choose conductivity.
Degassed conductivity
Connected conductivity
sensor
CO2 concentration
Read only
View this value in a user-defined measuring
screen or output the value via the current output.
Mathematical function
assignment view
Overview of the configured functions
Dual conductivity
You can subtract two conductivity values and use the result to monitor the efficiency of an
ion exchanger, for example.
Menu/Setup/Additional functions/Mathematical functions/MF1 to 6/Mode = Dual conductivity
Function
Options
Info
Calculation
Selection
• Off
• On
Factory setting
Off
On/off switch for the function
Inlet
The options depend on the
sensors connected
Select the sensors that should be used as the
minuend (
Inlet
, e.g. sensor upstream from the
ion exchanger) or subtrahend (
Outlet
, e.g.
sensor downstream from the ion exchanger).
Measured value
Outlet
Measured value
Main value format
Selection
• Auto
• #
• #.#
• #.##
• #.###
Factory setting
Auto
Specify the number of decimal places.
Cond. unit
Selection
• Auto
• µS/cm
• mS/cm
• S/cm
• µS/m
• mS/m
• S/m
Factory setting
Auto
Dual conductivity
Read only
View this value in a user-defined measuring
screen or output the value via the current output.
Mathematical function
assignment view
Overview of the configured functions