Electrical connection
Hauser
31
2. To fold out the carrier board towards the front, proceed as described in the "Routing
the cables" section .
3. Release a suitable hose gland on the right-hand underside of the analyzer and
remove the dummy plug from the gland.
4. Guide the spiral hose through the hose gland.
5.
Order versions with G' and NPT cable glands:
replace the pre-mounted M
‐
thread
cable glands with the G' or NPT glands enclosed. This does not affect the M32 hose
glands.
6. Comply with the instructions for protective grounding when installing the analyzer.
7. Remove the protective cover in the upper right-hand corner.
A0028925
8. Route the cable in the housing in such a way that the
exposed
cable shield fits into
one of the cable clamps and the cable cores can be easily routed as far as the plug-in
terminals.
9. Open the cable clamp and clamp the cable in place. Then tighten the screw of the
cable clamp.
10. Connect the cable or cables (depending on the version) to the following plug-in
terminals:
SP2
N
L
PE
86
85
97
98
SP1
N
L
PE
86
85
97
98
BK
BK
GN/YE
GY
PK
GN
YE
1
2
1
2
A0028924
33
Connection scheme for Liquiline System CAT820 / 860
1
Hose heater 100 to 120 V/200 to 240 V AC (optional)
2
Connections for Memosens and communication with analyzer (optional)
11. Secure the protective cover after connecting. Make sure that no cables or hoses are
jammed.
12. Use the 6 screws to secure the carrier plate after connecting.