Liquiline System CA80NO
Operation
Hauser
55
Icon
Location
Description
At measured value
The measured value is influenced by a simulated value
At measured value
The displayed measured value is simulated (for sensors)
Header bar
Controller is active
1)
Only pH or ORP measurement
If two or more diagnostic messages occur simultaneously, only the icon for the
message with the highest priority is shown on the display (for the order of priority
according to NAMUR, → 119).
10.1.4 Assignment views
Assignment views, e.g.
Channel assignment view
, appear as the last function in many
sections of the menu. You can use this function to see which actuators or functions are
connected to an input or output. The assignments appear in hierarchical order.
10.2 General settings
10.2.1 Basic settings
Menu/Setup/General settings
Function
Options
Info
Temperature unit
Selection
• °C
• °F
• K
Factory setting
°C
Current output range
Selection
• 0..20 mA
• 4..20 mA
Factory setting
4..20 mA
In accordance with Namur NE43, the linear
range is from 3.8 to 20.5 mA (
4..20 mA
) or
from 0 to 20.5 mA (
0..20 mA
). If the range is
exceeded or undershot, the current value stops
at the range limit and a diagnostic message (460
or 461) is output.
Failure current
0.0 to 23.0 mA
Factory setting
22.5 mA
The function meets NAMUR NE43.
‣
Set the current value that should be output
at the current outputs in the event of an
error.
The value for
Failure current
should be outside the measuring range. If you decided that your
Current
output range
=
0..20 mA
you should set an error current between 20.1 and 23 mA. If the
Current
output range
=
4..20 mA
you could also define a value < 4 mA as the error current.
The device allows an error current within the measuring range. In such instances pay attention to any
effects this may have on your process.
Alarm delay
0 to 9999 s
Factory setting
0 s
The software displays only the errors that are
present longer than the set delay time. This
makes it possible to suppress messages that only
occur briefly and are caused by normal process-
specific fluctuations.
Device hold
Selection
• Disabled
• Enabled
Factory setting
Disabled
You can enable an immediate, general hold (for
sensors) here. The function acts in the same way
as the
HOLD
soft key in the screens.