15.9 Process
Sample temperature
4 to 40 °C (39 to 104 °F)
Consistency of the sample
Low solids content (turbidity < 50 NTU), aqueous, homogenized
Sample supply
Unpressurized
15.10 Mechanical construction
Dimensions
Weight
Order version
Weight with cooling module
Weight without cooling module
Cabinet version
42 kg (92.6 lbs)
39.5 kg (87.1 lbs)
Open installation 34 kg (74.96 lbs)
31.5 kg (69.45 lbs)
Analyzer stand
75 kg (165.3 lbs)
72.5 kg (159.8 lbs)
Materials
Parts not in contact with medium
Cabinet version, exterior cover
Plastic ASA+PC
Open installation, exterior cover
Cabinet version, interior lining
Plastic PP
Open installation, interior lining
Window
Shatterproof glass, coated
Reagent container
Plastic PP
Insulation
Plastic EPP (extruded PP)
Base, analyzer stand
Powder-coated sheet steel
Parts in contact with medium
Dispensers
Plastic PP and elastomer TPE
Liquid Manager
Plastic PP and elastomer FKM
Hoses
C
‐
Flex, NORPRENE
Optical window
Glass
Molded seal
Elastomer EPDM
Drain pipe
Plastic PP
Sample collecting vessel (optional)
• Beaker
• Cover
• Level detector pins
• Seal
• Plastic PMMA
• Plastic PP
• Stainless steel 1.4404 (V4A)
• EPDM
Valve (optional)
PVDF