Liquiline System CA80NO
Commissioning
Hauser
51
2
3
RK
C
S1
P
RK
S1
C
W1
Cu
M
D
D5
D6
D7
optional
cooling
7
P
1
W2
optional
dilution
A0033656
69
Liquiline System CA80NO, self-priming
C
Cleaner
Cu
Cuvette
S1
Standard 1
P
Sample
RK
Reagent RK
2, 3, 7
Dispensers
W1
Outlet
D
Outlet
M
Photometer/measuring cell
1
T-piece
9.1.2
Connecting the sample inlet hose
1. Ensure a constant and sufficient supply of sample at the installation location.
2. Connect the liquid-bearing hoses of the sample supply system.
3. Self-priming system: connect the intake hose supplied (1.5 m (4.92 ft)) to the Liquid
Manager ("sample") (→ 49) and guide it through the hose gland of the analyzer to
the outside.
4. If present, connect the communication cable and hose heater of the sample
preparation system to the analyzer.
5. Ensure that only sample that has a low solids content is supplied, as otherwise there
is a risk of blockage.
9.2
Function check
L
WARNING
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions!
‣
Check that all connections have been established correctly in accordance with the
wiring diagram.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.
Before putting the device into operation:
1. Connect the liquid-bearing hoses of the sample supply system.
2. If present, connect the communication cable and hose heater of the sample
preparation system to the analyzer.
3. After mounting:
Check all connections to ensure they are secure and leak-tight.
4. Check that the hoses of the sample preparation system are mounted correctly in the
hose glands. It should not be possible to remove the hoses without the application of
force.