![Advanced Instruments OsmoPRO Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/advanced-instruments/osmopro/osmopro_user-manual_2856041090.webp)
Appendix E:
OsmoPRO MAX LIS Specification
90
OsmoPRO MAX Automated Osmometer User Guide
Example
O|1|3MA005||^^^||20161110082002|
20161110082002||0.000|||||
20161110082002||||||||20161110082002|
||<CR><ETX>
Result Record
The fields used in this record are as specified in LIS2-A2:
9.1
Record type ID—
R
identifies the record as a result
record.
9.2
Sequence number—Sequence number (Integer)
of the result record associated with the preceding
test order record.
9.3
Universal test ID—Text field containing up to four
parts describing the test. This code is always the
four-letter code
OSMO
for the instrument.
9.4
Measured value—Number (Integer) indicating the
osmolality value.
9.5
Units—Text field indicating the units for the
preceding value. This is always
mOsm/Kg H2O
for the instrument.
9.9
Result status (optional)—Text field containing the
status of the test. See the LIS2-A2 document for
the possible codes.
9.11
Operator identification (optional, UTF8-
encoded)—Text field that identifies the operator
who performed the test and possibly also the test
verifier, separated by a component delimiter.
9.12
Date/time of start of test—Text field containing
the date/time the test was started.
9.13
Date/time of end of test—Text field containing
the date/time the test finished.
9.14
Instrument Identification—Text field containing
the instrument serial number.
Example
R|1|^^^OSMO|51|mOsm/Kg H2O||N|N|F|
|OperatorID|20161027142723|
|17010095A<CR><ETX>
Request Information Record
The instrument can use this record to request the order
for a sample if it is not already present.
The LIS can use it to request results from previously
measured samples.
The fields used in this record are as specified in LIS2-A2:
11.1
Record type ID—
Q
identifies the record as a
request record.
11.2
Sequence number—Sequence number (integer)
of the request.
11.3
Starting range ID—Text field containing the
starting record to look up. It can have up to three
components:
The first must be the sample ID (see
)
or blank
The second must be the specimen ID (see
sections
A wildcard character of
*
can be used for both
sample ID and specimen ID to indicate partial
matches.
Either a sample ID or a specimen ID must be
indicated (wcard only is fine).
11.5
Universal test ID—Text field containing up to four
parts describing the test type requested. This
code is always the four-letter code
OSMO
for the
instrument.
11.6
Nature of request time limits (optional)—Text field
containing the base for the time request
described below.
R
(default) indicates the result test date.
S
indicates the sample collection date.
11.7
Beginning request results date and time
(optional)—Text field containing the oldest date/
time to use in the look up. Defaults to all times if
date/time indicated.
11.8
Ending request results date and time (optional)—
Text field containing the newest date/time to use
in the look up.
11.9
Requesting physician name (optional)—Text field
containing the name of the doctor requesting the
result.
11.10
Requesting physician telephone number
(optional)—Text field containing the telephone
number of the doctor requesting the result.
11.11
User field (optional)—Text field containing any
special instructions for the request.