Chapter 10:
Troubleshooting
OsmoPRO MAX Automated Osmometer User Guide
78
Displayed/printed
message
Error
Code
Type/
Severity
Possible cause/Suggested action
Fan error
96
Cause:
Cable not connected
Cause:
Bad fan
Action:
Check cable
If cable connection is good and problem persists after reboot, contact AI Service.
B scanner error
97
Cause:
Cable (RX or TX)
Action:
Check cable
Cause:
Bad back barcode scanner
Action:
Reboot
If cable connection is good and problem persists after reboot, contact AI Service.
B scanner not connected
98
Cause:
Cable
Action:
Reboot
Action:
Check cable
If cable connection is good and problem persists after reboot, contact AI Service.
F scanner error
99
Cause:
Cable (RX or TX)
Action:
Check cable
Cause:
Bad front barcode scanner
Action:
Reboot
If cable connection is good and problem persists after reboot, contact AI Service.
F scanner not connected
100
Cause:
Cable
Action:
Reboot
Action:
Check cable
If cable connection is good and problem persists after reboot, contact AI Service.
Assurance tag reader error 101
Cause:
Cable (RX or TX)
Action:
Check cable
Cause:
Bad assurance tag reader
Action:
Reboot
If cable connection is good and problem persists after reboot, contact AI Service.
Assurance tag reader not
connected
102
Cause:
Cable (RX or TX)
Action:
Check cable and reboot
If cable connection is good and problem persists after reboot, contact AI Service.
Liquid detection not
connected
103
Cause:
Cable issue or bent pipette.
Action:
Check cable and reboot
If cable connection is good and problem persists after reboot, contact AI Service.
Liquid detection error
104
Cause:
Cable Issue, bent pipette, or defective integrated circuit
Action:
Check cable and reboot
If cable connection is good and problem persists after reboot, contact AI Service.
Pipette error
105
Cause:
Pipette crashed
Action:
Check for pipette crash
If pipette crashed, contact AI Service.
Low cleaning fluid
106
Cause:
Too many power cycles.
Cause:
Too many back-to-back purge cycles
Action:
Load a new OsmoPRO MAX Maintenance Kit (see