OsmoPRO MAX Automated Osmometer User Guide
82
Sample types
Calibration standards, reference solutions, controls, plasma, serum, and urine
Units of Measure
mOsm/kg H
2
O
Measurement range
0 to 2000 mOsm/kg H
2
O
Display Range
0 to 2200 mOsm/kg H
2
O
Minimum required sample
volume
170
μ
L
Sample capacity
Continuous loading of up to 500 samples
Test time
~ 3 minutes (typical) to test a single sample
~ 1 hour (typical) to test 20 samples
Calibrated range
0 to 2000 mOsm/kg H
2
O
Calibration points
4 points (0, 300, 850, and 2000 mOsm)
Power requirement
100 to 240 VAC ±10% (50/60 Hz)/1.0 A
Power consumption: 60 W max.
Fuses (2): 250 V time delay (Type T) 2.0-Amp
Memory capacity
1,000 results/10,000 events
Memory backup
Integral lithium cell; 10-year life (unpowered); not user-replaceable
Classifications
Protection class: I
Over-voltage category: II
Pollution degree: 2
Moisture protection: IPX0 (ordinary)
Communication
USB 2.0 Type A ports (3)
USB 2.0 Type B port (1)
Ethernet 10/100
Operating conditions
Temperature: 18 °C to 35 °C (64 °F to 95 °F)
Humidity: 5 to 80% relative humidity (non-condensing)
Maximum operating altitude: 2000 m
Storage temperature
-40 °C to 45 °C (-40 °F to 113 °F)
Dimensions (D x W x H)
14” (36 cm) W x 16.5” (42 cm) D x 18” (46 cm) H including containers for fluids and
excluding the power cord, printer, and any USB peripherals
Net weight
13.6 kg (30 lb)
Shipping weight
20.9 kg (46 lb)
Warranty
One-year limited warranty on workmanship and parts
Certifications
Appendix B:
Product Specifications