Chapter 9:
Diagnostics
OsmoPRO MAX Automated Osmometer User Guide
70
9.4 Testing the Sample Probe
The Sample Probe function tests the sample probe at
selected positions on the turntable.
1.
Place a sample of deionized water in each sample
position on the turntable that you want to test.
2.
From the Home screen, press
Diagnostics > Sample
Probe
.
3.
Select a
Position
to test.
4.
Press
Load
.
5.
Press
Test
.
6.
Repeat steps 3 through 5 to test other sample
positions.
7.
Press
Exit
to return to the Diagnostics menu.
9.5 Testing the Barcode Scanners
1.
From the Home screen, press
Diagnostics > Barcode
.
2.
Place a loaded tube adapter onto the turntable.
3.
Manually move the turntable so the tube adapter is
located in front of the barcode scanner..
4.
If successful, you hear a beep and the barcode
information appears.
CAUTION:
The front barcode
scanner uses a Class 2 laser.
Do not stare directly into the
beam when scanning barcodes.
5.
Scan the barcode of an OsmoPRO MAX Maintenance
Kit, Calibration Kit, or calibration standard on the
front barcode reader.
6.
If successful, you hear a beep and the barcode
information appears.
7.
Press
Exit
to return to the Diagnostics menu.
The front barcode scanner supports the following
barcode types:
•
Code 39
•
Code 128
•
ECC
•
UPC
•
GS1
•
Codabar
•
Code 93
•
Code 11
•
Interleaved
•
PDF417
•
MSI
•
QR
•
IATA
•
MaxiCode
•
Aztec
•
EAN
9.6 Viewing and Exporting Instrument Events
An
event
is an action that occurs with the instrument.
The OsmoPRO MAX records the following events in the
events database for operators to review and export as
needed.
•
STAT tests
•
System alerts/alarms
•
Settings changes (users added/deleted, controls
added/deleted
•
Power up/power down
•
Calibration (see
Running Instrument Calibration
•
User login/logout
•
Data management (data import/export/backup - see
•
Instrument cleaning (see
•
Maintenance Kit replacement (
Instrument and
Consumables Maintenance
)