Chapter 2:
Installation and Setup
OsmoPRO MAX Automated Osmometer User Guide
20
Remove the Rear Panel
You must remove the rear panel of the instrument to
access the shipping stop.
Required
: Medium-sized Phillips screwdriver
1.
Position the instrument so that the serial number
plate
[A]
faces you.
2.
Remove the four circled Phillips head screws with the
screwdriver and save for later re-installation.
A
3.
Slide the rear panel down slightly and then pull it
away from the instrument.
Remove the Shipping Stop
Required
: Medium-sized Phillips screwdriver
The shipping stop
[B]
is shown below. It is secured with
a Philips head screw
[C]
.
B
1.
Unscrew the Philips head screw.
2.
Gently push the shipping stop forward approximately
1” (2.5 cm) to gain clearance.
3.
Rotate the shipping stop to the left so that it clears
the linear slide.
4.
Once it is clear from the slide, pull the shipping stop
back and out of the device (avoid wiring).
5.
Store the screw and shipping stop in a safe location in
case you need to use it for return shipping.
C