Chapter 2:
Installation and Setup
OsmoPRO MAX Automated Osmometer User Guide
26
5.
Hold the OsmoPRO MAX Maintenance Kit assurance
tag
[A]
against the assurance tag reader
[B]
until the
instrument sounds a tone.
A
B
The tone indicates that the kit is loaded in the system and
the test count updates to 500.
6.
Press
Next
.
Connect the Waste Bottle
1.
Connect the cover to the new waste bottle.
NOTE:
For subsequent waste fluid disposals, make sure
to dispose of the waste fluid according to
established laboratory procedures and to
disinfect the waste bottle using 1% hypochlorite
solution
2
after emptying the it.
2 Commercial bleach with 8.25% sodium hypochlorite diluted to 1% with water
2.
Position the waste bottle in the rear recess to the left
of the turntable.
3.
Press
Next
.
Install the Peristaltic Tubing Lines
Complete the following steps to install the tubing line for
each peristaltic pump.
1.
Press
to view the onscreen instructions.
2.
Lift the lever
[C]
to the upper position and remove
the cover.
C
3.
Insert the tubing clips
[D]
into the recesses of the
cover with the open end of the tubing clips pointing
out so that the tube sits symmetrically in the cover.
4.
Push the tubing into the recess with the tubing
touching the upper portion of the cover.