Chapter 4:
Using the Controls Log Database
OsmoPRO MAX Automated Osmometer User Guide
43
4.2 Deleting a Control
1.
Press
> Controls Log
and log in.
2.
Press to select the control to delete.
3.
Press
Delete
.
4.3 Managing Control Lots
For each control you add to the system, you can add,
delete, and select the default lot for the system to use
when testing.
Adding a Control Lot
To add a lot to a control:
1.
Press
> Controls Log
and log in.
2.
Press to select a control on the Controls Log screen.
3.
Press
Lots
.
4.
Press
Add
.
5.
Scan in the UDI barcode using the front barcode
scanner by aligning the barcode with the barcode
laser.
CAUTION:
The front barcode
scanner uses a Class 2 laser.
Do not stare directly into the
beam when scanning barcodes.