Chapter 2:
Installation and Setup
OsmoPRO MAX Automated Osmometer User Guide
29
Configure Operator/Testing Access, Quality Control, and
System Timeouts (Supervisors only)
Supervisors can perform the following operations from
the
Settings > Configuration
screen:
•
Configure permissions to certain functions for
Operator-level users.
•
Configure log in, sample ID, and replicate
requirements when testing samples.
•
Configure requirements for quality control testing
frequency using Clinitrol™.
•
Enable/disable operator login using badge barcode
scanning.
Configure Operator Permissions
To change Operator permissions for functions, complete
the following steps.
1.
Press
> Settings > Configuration
and log in.
2.
Select/deselect checkboxes on the
Operator
Permissions
tab to enable/disable operator access to
the following features:
–
–
Set Up Networking Communications
–
–
Running Instrument Calibration
–
Logging Routine Instrument Cleaning
–
–
Using the Controls Log Database
3.
Press
Save
.
4.
Press
Exit
to return to the Settings menu.
Configure Testing Access and Replicates
Authorized users can configure the system to require
users to enter a user ID and/or a sample ID before
testing can start.
1.
Press
> Settings >H/W Options
and log in.
Settings > Configuration > Testing
.
2.
Select the
User ID
checkbox to require users to enter
a user ID before testing.
3.
Select the
Sample ID
checkbox to require users to
enter a sample ID before testing.
4.
Select the number of replicates
Per Sample
(
1
through
10
).
5.
Select the number of replicates
Per Control
(
1
through
10
).
6.
Press
Save
.
7.
Press
Exit
to return to the Settings menu.
Configure Quality Control Actions
Supervisors can configure the system to perform one of
the following actions when a sample result is outside the
control limits:
•
Perform no action
•
Issue a warning but continue with the test run
•
Stop the entire test run
They can also configure the system to lock the
instrument if testing with Clinitrol 290 has not been
performed for a certain amount of time.
Configured actions only work with controls that are
entered in the controls database (see
) and if sample IDs are required during
testing (see
Configure Testing Access and Replicates
).