Chapter 2:
Installation and Setup
OsmoPRO MAX Automated Osmometer User Guide
25
Set the Installation Date
If the installation date of the instrument has not been
entered into the system, the following message appears
the first time an instrument powers up:
This message reappears every hour until the installation
date is set.
Complete the following steps to enter the installation
date to disable this message and enable the preventative
maintenance reminders based on this date.
1.
From the Home screen, press
to open the Main menu.
2.
Press
Diagnostics > Maintenance
and log in to the
Admin
account (see
).
3.
Press the
Installation
tab.
4.
Press
and select an
Installation Date
(required)
from the calendar.
Make sure
you select the correct date. You cannot
change the installation date once it is saved.
5.
Press
OK
.
6.
Press
Edit
to type a
Comment
(required - up to 1024
characters) to enter the operator who set the installation
date and any other information pertaining to the installation.
7.
Press
Save
.
8.
Press
Exit
to return to the Diagnostics menu.
9.
Press
to return to the Home screen.
Load the Maintenance Kit
Required materials
OsmoPRO MAX Maintenance Kit
(p/n MAX500) consisting of:
•
OsmoPRO MAX System
Fluid
•
OsmoPRO MAX
Enzymatic Cleaner
•
OsmoPRO MAX
Replacement Tubing
Lines (for peristaltic
pump)
1.
From the Home screen, press
to open the Main
menu.
2.
Select
Maintenance Kit
from the Main menu.
3.
Log in to the
Admin
account (see
) and follow the instructions on
the
Maintenance Kit
screen (also provided below).
4.
Press
Load
.