Chapter 4:
Using the Controls Log Database
OsmoPRO MAX Automated Osmometer User Guide
45
Changing the Default Control Lot
To change the default control lot used in testing:
1.
Press
> Controls Log
and log in.
2.
Press the control you want to change.
3.
Press
Lots
.
4.
Select the control lot to set as default.
Y
5.
Press
Default
.
6.
Press
Exit
to return to the Controls Log screen.
4.4 Displaying and Exporting a Levey-
Jennings Chart
Your instrument can generate Levey-Jennings charts for
controls with standard deviation (SD) bands.
To generate a chart, the control must be entered in the
controls database.
1.
Press to select a control on the Controls Log screen.
2.
Press
Chart
.
3.
Select
1
,
2
, or
3
SD Bands.
4.
Select a date range using the
Min
and
Max
calendars
(
).
5.
Press
Export
to export the chart image to a USB
drive.
6.
When you have finished viewing and/or exporting the
chart, press
Exit
to close the chart and return to the
Controls screen.