Chapter 2:
Installation and Setup
OsmoPRO MAX Automated Osmometer User Guide
28
Clean the Fluidics
1.
Place the Enzymatic Cleaner adapter in position 20
on the turntable.
2.
Unscrew the cap on the Enzymatic Cleaner vial.
3.
Place the Enzymatic Cleaner vial in the adapter.
4.
Press
Clean
.
The cleaning cycle takes approximately 10 minutes to
complete.
5.
When the fluidics cleaning is complete, remove the
adapter and vial from the turntable.
6.
Dispose of the vial according to the disposal
instructions in the Safety Data Sheet.
7.
Press
Done
to return to the Home screen.
The turntable is unlocked and the system fluid indicator
on the Home screen indicates that fluid is now available
to test 500 samples.