
Chapter 10:
Troubleshooting
OsmoPRO MAX Automated Osmometer User Guide
77
Hardware errors
Osmometer Board
Displayed/printed
message
Error
Code
Type/
Severity
Possible cause/Suggested action
Temperature and
humidity not detected
4
Cause:
Issue reading and writing to sensor
Action:
Reboot instrument
If problem persists after reboot, contact AI Service.
Power supply error,
osmometer board
5
Cause:
Error on 3.3 V ADC sensor
Action:
Reboot instrument
If problem persists after reboot, contact AI Service.
Cooling engine setup
required
43
Cause:
Instrument needs module setup
Contact AI Service.
Calibration is required
44
Cause:
Instrument needs calibration
Action:
Calibrate the instrument
Motor Board
Displayed/printed
message
Error
Code
Type/
Severity
Possible cause/Suggested action
Temperature and humidity
not detected.
68
Cause:
Issue reading and writing to sensor
Action:
Reboot instrument
If problem persists after reboot, contact AI Service.
Power supply error, motor
board
69
Cause:
Error on 3.3 V ADC sensor
Action:
Reboot instrument
If problem persists after reboot, contact AI Service.
Comm error
92
Cause:
Cable issue
Action:
Check cable
Cause:
Bad scanner
Action:
Reboot instrument
If cable connection is good and problem persists after reboot, contact AI Service.
Comm error
93
Cause:
Cable issue
Action:
Check cable
Cause:
Bad assurance tag reader
Action:
Reboot instrument
If cable connection is good and problem persists after reboot, contact AI Service.
EEPROM error
94
Cause:
Occasional issue with new boards
Action:
Reboot instrument.
If problem persists after reboot, contact AI Service.
EEPROM error
95
Cause:
Occasional issue with new boards
Cause:
Power not in range
Action:
Reboot instrument
If problem persists after reboot, contact AI Service.