11-60 ADS
Manual
Upward Depth Subsystem
The following tables contain general troubleshooting techniques for
the upward depth subsystem corresponding to the AV|Gated,
AV|Max and Peak Combo sensor.
Note:
Contact your regional ADS representative for
further diagnosis prior to replacing a sensor or monitor.
Problem
UpDepth data periodically reads 0.88 inches (22 mm).
Possible
Causes
Minimum flows are occurring within sensor deadband
(extending 1.0 inches (25 mm) up from bottom of Peak
Combo Sensor) or the sensor is receiving a strong
return signal close to the sensor face.
Sensor transmit/receive surfaces are dirty.
Sensor or cabling might be faulty.
Possible
Solutions
Verify minimum flow depths are not occurring within
sensor deadband.
Set the deadband on the UpDepth sensor to a setting
other than
Auto
, but well below the expected minimum
flow depth. Fire the sensor several times to verify
whether the readings improve.
Ensure the sensor is installed as closely as possible to
the 6:00 position in the pipe, silt allowing.
Clean the sensor transmit/receive surfaces.
Inspect the sensor cables for integrity.
Check for moisture in the connector.
Contact your regional ADS representative.
Replace the sensor.
Replace the monitor if defective.
Содержание 507165
Страница 20: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Страница 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Страница 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Страница 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Страница 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Страница 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Страница 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Страница 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Страница 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Страница 53: ...System Overview 2 13 ExPAC XBUS ...
Страница 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Страница 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Страница 292: ...5 86 ADS TRITON Manual Confirm Level Front to Back Installed ParaDepth ParaFlow Wall Mount Bracket ...
Страница 315: ...External Power 6 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 327: ...External Power 6 21 Wires exiting DC output terminal on power supply POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS ...
Страница 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Страница 358: ...Inclinometer 8 3 ...
Страница 507: ...C 4 ADS TRITON Manual ...
Страница 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...