1-8 ADS
Manual
(ExPAC), the Monitor Assembly shall be bonded to the
ExPAC earth with a conductor having a minimum cross-
sectional area of 4 mm
2
, to achieve a resistance
≤
1
Ω
.
Other Conditions for Safe Use
z
The ADS Model 8000-FST-IM shall use only the 12-volt IS
Battery Pack (ADS p/n 8000-0043) for internal power.
z
Only ADS-approved sensors shall be connected to the ADS
Model 8000-FST-IM: ADS sensor types CSX series.
z
The only communication devices that may be connected to the
COMM
+ EXT PWR
connector on the
are the
USB Serial Interface (ADS p/n 8000-0337), the
ExPAC/XBUS/XIO power cable (ADS p/n 8000-378-10/25),
the Direct Connect Interface (ADS p/n 8000-0054), the
Inclinometer (ADS p/n 8000-0528), and the Sampler Cable
(ADS p/n 8000-0348).
Note:
The Triton-Telog Comm Cable (ADS p/n 8000-
0054-01) also may be connected to the
COMM + EXT
PWR
connector on the
to provide
communication between the Telog Ru-33 or Ru-35 and the
monitor. However, because the Ru-33 nor Ru-35 are not
certified for IS operation, an installation involving a
connection between these units will not be considered
intrinsically safe, unless the Telog unit is installed outside
the hazardous area.
Note:
The Ayyeka Wavelet 4R Comm Cable (ADS p/n
8000-0054-03) also may be connected to the
COMM +
EXT PWR
connector on the
to provide
communication between the Ayyeka Wavelet 4R monitor
and the
. However, because the Wavelet 4R is
not certified for IS operation, an installation involving a
connection between this unit will not be considered
intrinsically safe, unless the Ayyeka unit is installed
outside the hazardous area.
Содержание 507165
Страница 20: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Страница 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Страница 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Страница 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Страница 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Страница 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Страница 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Страница 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Страница 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Страница 53: ...System Overview 2 13 ExPAC XBUS ...
Страница 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Страница 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Страница 292: ...5 86 ADS TRITON Manual Confirm Level Front to Back Installed ParaDepth ParaFlow Wall Mount Bracket ...
Страница 315: ...External Power 6 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 327: ...External Power 6 21 Wires exiting DC output terminal on power supply POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS ...
Страница 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Страница 358: ...Inclinometer 8 3 ...
Страница 507: ...C 4 ADS TRITON Manual ...
Страница 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...