Sensor Installation and Connection 5-29
3.
Place the ring in the input pipe
at least
12 inches (305 mm)
upstream from the manhole or edge of the pipe with the
sensors facing upstream toward the oncoming flow. It must
be located far enough upstream from the manhole to minimize
the effect of the draw-down caused by a possible drop in the
manhole invert.
Installing the ring at least 12 inches (305 mm) upstream from the manhole
invert
Keep the following in mind:
The installed AV|Gated sensor must be positioned in the
bottom, center of the pipe and
cannot /- 5
degrees from the 6:00 position.
If a greater rotation
/- 5 degrees is needed, then a Silt Mount Adapter
must be used. Install the AV|Gated sensor and then verify
the sensor is correctly positioned. After activating the
monitor, perform a Diagnostic reading on the AV|Gated
Ultrasonic sensor and verify the
Sensor Slope
and
Sensor
Rotation
values are not greater than 5.0. See
Chapter 4,
Running Sensor Diagnostics
for more information.
Be sure to mount the ParaDepth, ParaFlow, Surface
Combo sensor/Ultrasonic Depth Sensor at the top (crown)
of the pipe and to mount the AV|Gated, AV|Max, and Peak
Combo sensors as close as possible to the bottom center of
the pipe,
above
any silt present and
below
the flow surface
(during minimum flows).
FLOW
12"
Содержание 507165
Страница 20: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Страница 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Страница 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Страница 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Страница 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Страница 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Страница 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Страница 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Страница 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Страница 53: ...System Overview 2 13 ExPAC XBUS ...
Страница 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Страница 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Страница 292: ...5 86 ADS TRITON Manual Confirm Level Front to Back Installed ParaDepth ParaFlow Wall Mount Bracket ...
Страница 315: ...External Power 6 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 327: ...External Power 6 21 Wires exiting DC output terminal on power supply POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS ...
Страница 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Страница 358: ...Inclinometer 8 3 ...
Страница 507: ...C 4 ADS TRITON Manual ...
Страница 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...