System Overview
2-27
ParaFlow sensor measuring surface velocity
Surcharge Peak Velocity Sensor
The Surcharge Peak Velocity sensor resides in the angled, front
portion of the sensor in front of the ultrasonic depth parabolic
cavity. It measures the velocity of the flow during surcharge
conditions using the same methodology as the Peak Combo
Velocity Sensor. Therefore, it will measure velocity
only
when the
pipe is full.
Surcharge Pressure Depth Sensor
The surcharge pressure depth sensor in the ParaFlow Sensor
measures pressure depth using the same technique as the other
combo sensors, except from the top of the pipe. Since it can
provide an accurate depth only when submerged, the data from this
sensor is useful only under full-pipe or surcharge conditions. The
ParaFlow Sensor supports pressure measurement of 0 - 10 PSI (0 -
0.68 Bar.
Peak Combo Sensor
The Peak Combo Sensor with a standard 35-foot (10.6-m) or 100-
foot (30.5-m) cable (ADS p/n 8K-CS4-XX-35/1H) typically mounts
at the bottom of the pipe and measures ultrasonic depth, pressure
Содержание 507165
Страница 20: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 27: ...Introduction 1 19 CS6 LRD Sensor Certificate of Conformity ...
Страница 29: ...Introduction 1 21 CSA Certificate of Compliance Page 1 of 7 ...
Страница 30: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 2 of 7 ...
Страница 31: ...Introduction 1 23 CSA Certificate of Compliance Page 3 of 7 ...
Страница 32: ...1 24 ADS TRITON Manual CSA Certificate of Compliance Page 4 of 7 ...
Страница 33: ...Introduction 1 25 CSA Certificate of Compliance Page 5 of 7 ...
Страница 34: ...1 26 ADS TRITON Manual CSA Certificate of Compliance Page 6 of 7 ...
Страница 35: ...Introduction 1 27 CSA Certificate of Compliance Page 7 of 7 ...
Страница 37: ...Introduction 1 29 Sample TRITON Declaration of Conformity Page 2 ...
Страница 53: ...System Overview 2 13 ExPAC XBUS ...
Страница 266: ...5 60 ADS TRITON Manual z Installing the Submersion Sensor z Measuring the physical offset ...
Страница 288: ...5 82 ADS TRITON Manual Topside Wall Mount Bracket Topside Sensor Bracket ...
Страница 292: ...5 86 ADS TRITON Manual Confirm Level Front to Back Installed ParaDepth ParaFlow Wall Mount Bracket ...
Страница 315: ...External Power 6 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 327: ...External Power 6 21 Wires exiting DC output terminal on power supply POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS ...
Страница 345: ...7 2 ADS TRITON Manual use of this installation guide or the installation of any ADS equipment ...
Страница 358: ...Inclinometer 8 3 ...
Страница 507: ...C 4 ADS TRITON Manual ...
Страница 535: ...Index In 19 wiring to the monitor 6 29 XIO device parameters 4 84 ...