1.1 Methods
Notes:
1. This test uses an attenuated radiation method for the reading of FAU (Formazin
Attenuation Units). The results can not be used for USEPA reporting purposes, but may
be used for routine measurements. The attenuated radiation method is different from
the Nephelometric method.
2. The estimated detection limit is 20 FAU.
3. Collect water samples in clean plastic or glass bottles and analyse the water sample
as soon as possible. It is possible to store the sample at 4 °C for 48 hours. Before
measurement warm up the sample to the temperature at collection time. Temperature
differences between measurement and sample collection can effect the turbidity of
the sample.
4. Color interferes if light is absorbed at 530 nm. For strong colored water samples a
filtered portion of the sample can be used for zeroing instead of the deionized water.
5. Air bubbles interfere and can be removed using an ultrasonic bath.
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......