Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Notes:
1. Temperature of the sample should be 15°C – 25°C.
2. If Silica or Phosphate is present a yellow color is developed.
3. In this step any yellow color due to Phosphate is removed.
4. Interferences:
Substance
Interference
Iron
large amounts interfere
Phosphate
does not interfere at concentrations less than 50 mg/l PO
4
at 60 mg/l PO
4
the interference is approx. – 2%
at 75 mg/l PO
4
the interference is approx. – 11 %
Sulfide
interferes at all levels
Occasionally water samples contain forms of silica which reacts very slowly with
Molybdate. The nature of these forms is not known.
A pre-treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will
make these forms reactive to Molybdate (pre-treatment is given in “Standard Methods
for the Examination of Water and Wastewater” under “ Silica Digestion with Sodium
Bicarbonate”).
5. SiO
2
Si
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......