Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Suspended Solids
0 – 750 mg/l TSS
Sample preparation
Blend approx. 500 ml of the water sample in a blender
at high speed for 2 minutes.
1. Fill a clean vial (24 mm Ø) with
10 ml of the
deionized water,
close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber and empty
the vial completely.
5. Stir the blended water sample. Immediately rinse the
vial with the water sample and fill with
10 ml water
sample
.
6. Place the vial in the sample chamber making sure that
the marks are aligned.
7. Press
TEST
key.
The result is shown in the display in mg/l TSS (Total
Suspended Solids).
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
4
8
3
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......