Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
General:
Ortho-Phosphate ions react with the reagent to form an intense blue color (methods
320
,
323
,
324
,
325
and
326
).
Phosphate in organic and condensed inorganic forms (meta-, pyro- and polyphosphates)
must be converted to ortho-Phosphate ions before analysis.
Pretreatment of the sample with acid and heat provides the conditions for hydrolysis of the
condensed inorganic forms. Organically combined phosphates are converted to ortho-
Phosphate ions by heating with acid and persulfate.
The amount of organically combined phosphates can be calculated:
mg/l Phosphate, organic = mg/l Phosphate, total – mg/l Phosphate, acid hydrolysable
In methods
321
and
327
the ortho-Phosphate ions react with the Vanadate-molybdate-
reagent under acid conditions to form a yellow colored product.
Notes – only for tube tests and tests with powder packs:
323, 324, 325, 326
1. Application: for water, wastewater and seawater.
2. Highly buffered samples or samples with extreme pH values should be adjusted
between pH 2 and pH 10 before analysis (with 1 mol/l Hydrochloric acid or 1 mol/l
Sodium hydroxide).
3. Interferences:
Large amounts of turbidity may cause inconsistent results.
Interfering substance
Interference level:
Aluminium
greater than 200 mg/l
Arsenate
at any level
Chromium
greater than 100 mg/l
Copper
greater than 10 mg/l
Iron
greater than 100 mg/l
Nickel
greater than 300 mg/l
Silica (Silicium dioxide)
greater than 50 mg/l
Silicate
greater than 10 mg/l
Sulfide
at any level
Zinc
greater than 80 mg/l
Phosphate, ortho = Phosphorus, reactive
~
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......