Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Notes:
1. Extremely basic or acidic water samples should be adjusted with 0.5 mol/l (1 N) Sulfuric
acid solution or 1 mol/l (1 N) Sodium hydroxide solution to pH 7.
2. Interferences:
Interfering substance
Interference levels and treatments
Calcium
greater than 1000 mg/l CaCO
3
Iron
Interferes at all levels. Correct as follows:
a) determine the concentration of iron present in the
sample by performing a total Iron test
b) add the same iron concentration as determined
to the deionized water (step 1).
The interference will be blanked out successfully.
Magnesium
greater than 6000 mg/l CaCO
3
Nitrate
greater than 100 mg/l NO
3
-N
Nitrite
greater than 12 mg/l NO
2
-N
Phosphate
greater than 100 mg/l PO
4
-P
Sulfate
greater than 300 mg/l SO
4
Sulfide
intensifies the color
Glycine, Hydrazine,
Color, Turbidity
Less common interferences such as Hydrazine and
Glycine will cause intensified colors in the prepared
sample. Turbidity and color will give erroneous high
values. Samples with severe interferences require
distillation.
3. N
NH
4
NH
3
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......