Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
Delete stored results
Press [MODE], [Shift] + [3][4] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [0] key to retain the data sets
in memory.
After pressing key [Shift] + [1] the following
acknowledgment is displayed:
Press [ ] key to delete.
ATTENTION:
All stored test results are deleted
or cancel without deleting data by pressing [ESC] key.
Note:
All stored test results are deleted.
2.4.5 Calibration
Fluoride Calibration
Caution: See notes!
Press [MODE], [Shift] + [4][0] keys.
Confirm with [ ] key.
The display shows:
1. Fill a clean vial (24 mm Ø) with exactly
10 ml
of deionised water
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
•
•
<Delete data>
Delete all data?
YES : 1 NO : 0
<Calibration>
170 Fluoride
Zero: deionized water
press ZERO
<Delete data>
Delete data
Do not delete: ESC
0
Mode
0
4
Mode
4
3
1
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......