Correct filling of the vial:
6. Always perform zeroing and test with closed vial cap.
7. Bubbles on the inside wall of the vial lead to incorrect measurements. To prevent this,
remove the bubbles by swirling the vial before performing the test.
8. Avoid spillage of water in the sample chamber. If water should leak into the instrument
housing, it can destroy electronic components and cause corrosion.
9. Contamination of the lens in the sample chamber can result in errors. Check at regular
intervals and – if necessary – clean the light entry surfaces of the sample chamber using
a moist cloth or cotton buds.
10. Large temperature differences between the instrument and the environment can lead
to errors – e.g. due to the formation of condensation in the area of the lens or on
the vial.
11. To avoid errors caused by stray light do not use the instrument in bright sunlight.
Insertion of the adapter:
adapter 16 mm Ø
adapter 13 mm Ø
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......