Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Table 3:
Interfering substances
Interference level using 5 ml of sample
Aluminum
100 mg/l
Arsenate
interferes at all concentrations
Benzotriazole
10 mg/l
Bicarbonate
1000 mg/l
Bromide
100 mg/l
Calcium
5000 mg/l
CDTA
100 mg/l
Chloride
5000 mg/l
Chromate
100 mg/l
Copper
100 mg/l
Cyanide
100 mg/l;
increase the UV digestion to 30 minutes
Diethanoldithiocarbamate
50 mg/l
EDTA
100 mg/l
Iron
200 mg/l
Nitrate
200 mg/l
NTA
250 mg/l
Orthophosphate
15 mg/l
Phosphite and
organophosphorus compounds
reacts quantitatively;
Meta- and Polyphosphates do not interfere
Silica
500 mg/l
Silicate
100 mg/l
Sulfate
2000 mg/l
Sulfide
interferes at all concentrations
Sulfite
100 mg/l
Thiourea
10mg/l
highly buffered samples
or extreme sample pH
may exceed the buffering capacity of the
reagents and require sample pretreatment
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......