Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Phosphate, ortho
with Reaction Tube Test
0.06 – 5 mg/l PO
4
Insert the adapter for 16 mm Ø vials.
1. Open the white cap of one
tube PO
4
-P Dilution
and
add
5 ml of the water sample.
2. Place the vial in the sample chamber making sure that
the
l
marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add the contents of
one Phosphate Rgt. F10
Powder Pack
straight from the foil to the water sample
(Note 1).
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 10-15 sec., Note 2).
7. Place the vial in the sample chamber making sure that
the
l
marks are aligned.
8. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l ortho-
Phosphate.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
2:00
1
2
3
4
Ø 16 mm
2
4
3
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......