3
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Notes:
1. Application: Testing of residual corrosion inhibitors (Oxygen scavengers) in boiler feed
water or condensate.
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinse thoroughly with
deionized water.
3. Ideally temperature for full color development is 25°C ± 3 °C.
4. Volume should always be metered by using suitable pipette (class A).
5. Keep blank and sample dark during color development time. UV-light (sunlight) causes
high measurement results.
6. Interferences:
• Iron (II) interferes at all concentrations:
Repeat the test procedure but without adding the DEHA Rgt 2 solution. If the
displayed result is above 20 μg/l subtract this value from the DEHA test result.
• Substances which reduce Iron (III) interfere. Substances which complex iron strongly
may interfere also.
• Substances who may interfere when present in concentrations at:
Borate (as Na
2
B
4
O
7
)
500 mg/l
Cobalt
0.025 mg/l
Copper
8.0 mg/l
Hardness (as CaCO
3
)
1000 mg/l
Lignosulfonates
0.05 mg/l
Manganese
0.8 mg/l
Molybdenum
80 mg/l
Nickel
0.8 mg/l
Phosphate
10 mg/l
Phosphonates
10 mg/l
Sulfate
1000 mg/l
Zinc
50 mg/l
7. There is an option to change the unit from mg/l to μg/l.
The unit mg/l is rounded, e.g.: 25 μg/l = 0.025 mg/l display 0.03 mg/l.
mg/l
μg/l
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......