Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Hydrazine
with Ampoule
0.01 – 0.7 mg/l N
2
H
4
/ 10 – 700 μg/l N
2
H
4
Insert the adapter for 13 mm Ø vials.
1. Place the blank in the sample chamber. The blank is part
of the test kit.
2. Press
ZERO
key.
3. Remove the blank from the sample chamber.
4. Fill the sample container to the 25 ml mark with the
water sample.
5. Place one ampoule in the sample container. Snap the
tip by pressing the vial against the side of the sample
container. The ampoule breaks at the neck and the vial
fills automatically. A small volume of inert gas remains
in the ampoule.
6. Mix the contents of the ampoule by inverting it several
times, allowing the bubble to move from one end to
the other. Dry the outside of the vial.
7. Place the ampoule in the sample chamber.
8. Press
TEST
key.
Wait for a
reaction period of 10 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as Hydrazine.
Countdown
10:00
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
0
7
2
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......