Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Hardness, Calcium
with tablet reagent
50 – 900 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with
10 ml deionized water
.
2. Add
one CALCHECK tablet
straight from the foil to
the deionized water and crush the tablet using a clean
stirring rod.
3. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
4. Place the vial in the sample chamber making sure that
the marks are aligned.
5. Press
ZERO
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
6. Remove the vial from the sample chamber.
7. Add
2 ml of the water sample
to the prepared vial.
Caution: Vial is filled up to the top! (Note 4)
8. Close the vial tightly with the cap and swirl several times
(5x) to mix the contents.
9. Place the vial in the sample chamber making sure that
the marks are aligned.
10. Press
TEST
key.
The result is shown in the display as Calcium Hardness.
Zero accepted
prepare Test
press TEST
prepare Zero
press ZERO
Countdown
2:00
9
0
1
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......