Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
13. Place the vial (the blank) in the sample chamber
making sure that the marks are aligned.
14. Press
ZERO
key.
Wait for a
reaction period of 2 minutes
(Note 7).
After the reaction period is finished the measurement
starts automatically.
15. Remove the vial from the sample chamber.
16. Place the vial (the sample) in the sample chamber
making sure that the marks are aligned.
14. Press
TEST
key.
The result is shown in the display as mg/L PO
4
3-
.
To calculate the actual phosphonate concentration multiply the reading with the
corresponding dilution factor from table 1.
To calculate the active phosphonate concentration multiply the actual phosphonate
concentration using the appropriate factor from table 2.
Notes:
1. Rinse all glassware with 1:1 Hydrochloric acid first and then reinse with deionized water.
2. During UV digestion Phosphonates are convertd to Ortho-Phosphates.
This step is normally completed in 10 minutes. High organic loaded samples or a weak
lamp can cause incomplet phosphate conversion.
3. UV lamp must be purchased separately.
4. While the UV lamp is on UV safety goggles must be worn.
5. For handling of the UV lamp see manufacturer's manual.
Do not touch the surface of the UV lamp. Fingerprints will etch the glass.
Wipe the UV lamp with a soft and clean tissue between measurements.
6. The reagent does not dissolve completely.
7. The given reaction time of 2 minutes refers to a water sample temperature of more
than 15 °C. At a sample temperature of lower than 15 °C a reaction time of 4 minutes
is required.
Tables:
see next page
Zero accepted
prepare Test
press TEST
prepare Zero
press ZERO
Countdown
2:00
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......