Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Hydrazine
with Powder Reagent
0.05 – 0.5 mg/l N
2
H
4
/ 50 – 500 μg/l N
2
H
4
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
(Note 1, 2), close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
1 g HYDRAZINE test powder
(Note 3) to the
water sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents.
7. Press
[ ]
key.
Wait for a
reaction period of 10 minutes.
After the reaction period is finished proceed as follows:
8. The slight turbidity that occurs when the reagent is added
must be removed by filtration (Note 4).
9. Place the vial in the sample chamber making sure that
the marks are aligned.
10. Press
TEST
key.
The result is shown in the display as Hydrazine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
start:
0
5
2
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......