7
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
12. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
13. Place the vial in the sample chamber making sure that
the marks are aligned.
14. Press
TEST
key.
Wait for a
reaction period of 10 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Urea.
Notes:
1. The sample temperature should be between 20°C and 30°C.
2. Carry out the test at the latest one hour after sample taking.
3. The tablets must be added in the correct sequence.
4. Do not store reagent 1 below 10°C; granulation is possible.
Store reagent 2 (Urease) in the refrigerator at a temperature of 4°C to 8°C.
5. The AMMONIA No. 1 tablet will only dissolve completely after the AMMONIA No. 2
tablet has been added.
6. Ammonia and chloramines are also measured during urea measurement.
7. Before analysing seawater samples, a measuring spoon of Ammonia Conditioning
Powder must be added to the sample and swirled to dissolve before AMMONIA No. 1
tablet is added.
8. Fill the vial with drops of the same size by holding the bottle vertically and squeeze
slowly.
9.
Due to the reagent’s storage and shipping requirements, this reagent is not
available in many locations.
Countdown
10:00
Zero accepted
prepare Test
press TEST
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......