23
Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
User method list, switch all methods on
This mode function activates all methods. After switching on the instrument a scroll list of
all available methods is automatically shown in the display.
Press [MODE], [Shift] + [6][1] keys.
Confirm with [ ] key.
The display shows:
• Press [Shift] + [1] key to display all methods in the
method selection list.
• Press [Shift] + [0] key to keep the valid method
selection list.
The instrument goes back to mode menu automatically.
User method list, switch all methods off
The program structure requires that the method list must have at least one active (switched
on) method. For this reason the instrument activates one method automatically.
Press [MODE], [Shift] + [6][2] keys.
Confirm with [ ] key.
The display shows:
• Press [Shift] + [1] key to display only one method in
the method selection list.
• Press [Shift] + [0] key to keep the valid method
selection list.
The instrument goes back to mode menu automatically.
<Mlist all on>
switch on all
methods
YES: 1, NO: 0
<Mlist all off>
switch off all
methods
YES: 1, NO: 0
0
0
Mode
1
6
Mode
2
6
1
1
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......