Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
Initialise User Method System (Polynomials & Concentration)
Power loss will cause incoherent data. The user method system must be initialised with this
mode function to set it to a predefined state.
ATTENTION:
All stored user methods (polynomial & concentration) are deleted with initialisation.
Press [MODE], [Shift] + [6][9] keys.
Confirm with [ ] key.
The display shows:
Confirm with [ ] key.
The query is displayed:
• Press [Shift] + [1] key to start initialisation.
• Press [Shift] + [0] key to to cancel without initialisation.
The instrument goes back to mode menu automatically.
<User m. init>
Start:
Initialising?
YES: 1, NO: 0
0
Mode
9
6
1