Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Chlorine dioxide
in absence of Chlorine
with Tablet
0.05 – 11 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber and
empty it,
leaving a few drops remaining in the vial
.
5. Add
one DPD No. 1 tablet
straight from the foil and
crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press
TEST
key.
The result is shown in the display
as Chlorine dioxide in mg/l Chlorine,
or
as Chlorine dioxide in mg/l ClO
2
.
Notes:
See page 51.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
*,** mg/l ClO
2
*,** mg/l ClO
2
[Cl]
2
0
1
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......