Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide
1.1 Methods
Chlorine HR (Kl)
with Tablet
5 – 200 mg/l Cl
2
Insert the adapter for 16 mm Ø vials.
1. Fill a clean vial (16 mm Ø) with
8 ml of the water
sample
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are
l
aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
one CHLORINE HR (KI) tablet
straight from
the foil to the water sample and crush the tablet using
a clean stirring rod.
6. Add
one ACIDIFYING GP tablet
straight from the
foil to the same water sample and crush the tablet using a
clean stirring rod.
7. Close the vial tightly with the cap and swirl several times
until the tablets are dissolved.
8. Place the vial in the sample chamber making sure
that the marks are
l
aligned.
9. Press
TEST
key.
The result is shown in the display in mg/l Chlorine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Ø 16 mm
0
5
1
Summary of Contents for AQUAfast AQ3700
Page 1: ...Thermo Scientific Orion AQUAfast AQ3700 Colorimetry Meter User Guide ...
Page 8: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 9: ... Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 1 Methods ...
Page 143: ...141 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 206: ...204 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 207: ...205 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 2 Instrument Manual ...
Page 210: ...208 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 254: ...252 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide ...
Page 255: ...253 Thermo Scientific Orion AQUAfast AQ3700 Colorimeter User Guide Part 3 Enclosure ...
Page 263: ......